Maestro 3D Dental Scanner - Calibration

From Maestro3D
(Difference between revisions)
Jump to: navigation, search
(Cos'è la calibrazione?)
(Cos'è la calibrazione?)
Line 1: Line 1:
== Cos'è la calibrazione? ==
+
== What is calibration? ==
 
<div align="justify">   
 
<div align="justify">   
  
In a 3D scan system, calibrate a camera means extract the position of the ground plane in the chosen reference and a serial of parameters, necessary to the following elaboration.These parameters are called intrinsics and they are:
+
In a 3D scan system, calibrate a camera means extract the position of the ground plane in the chosen reference and a serial of parameters, necessary to the following elaboration. These parameters are called intrinsics and they are:
  
* focal length
+
* focal length (fo)
 
* image center (cc)
 
* image center (cc)
 
* radial distortion factor (k)
 
* radial distortion factor (k)
  
 
As told before, the goal of the camera calibration is to estimate the intrinsics parameters (focal length, image center and radial distortion factor) and to identify the position of the ground plane in the solidal reference with the camera.
 
As told before, the goal of the camera calibration is to estimate the intrinsics parameters (focal length, image center and radial distortion factor) and to identify the position of the ground plane in the solidal reference with the camera.
Come anticipato precedentemente, lo scopo della calibrazione della telecamera è quello di stimare i parametri intrinseci (lunghezza focale, centro dell’immagine e fattore di distorsione radiale) e di individuare la posizione del piano terra nel riferimento solidale con la telecamera.
 
  
La calibrazione delle telecamere è un'operazione essenziale per estrarre informazioni metriche da immagini 2D.  
+
Cameras calibration is a basic procedure to extract metric piece of information from 2D images.
Una telecamera è generalmente modellata mediante il modello proiettivo centrale pinhole camera (o foro stenopeico). Tale modello è definito da due famiglie di parametri:
+
A camera is usually represented by the central projector model pinhole camera. This model is defined by two parameters:
  
 +
* Intrinsic parameters (from three to five): they describe the camera independently from their position in the space.
 +
* Extrinsic parameters (six): they describe camera position in the space, independently from their internal features.
  
* Parametri intrinseci (da tre a cinque): descrivono la telecamera indipendentemente dalla sua posizione nello spazio.
+
This model must be conveniently correct to consider optical distortions introduced by camera lenses. The use of this kind of lenses, introduces a distortion which is necessary to balance before making a whatever calculation.
* Parametri estrinseci (sei): descrivono la posizione della telecamera nello spazio indipendentemente dalle sue caratteristiche interne.
+
  
  
Questo modello deve essere opportunamente corretto per tenere conto delle distorsioni ottiche introdotte dalle lenti della telecamera. L’uso di tali lenti introduce una distorsione che è necessario compensare prima di poter procedere ad una qualsiasi misurazione.
+
In literature there are several models to describe this kind of phenomenon. The model which has been studied and developed by Maestro 3D, estimates the lenses distortions in order to obtain a calibration which allows an excellent attention in the scanning process.  
  
  
In letteratura sono presenti diversi modelli per descrivere tale fenomeno. Il modello studiato e sviluppato da Maestro 3D stima le distorsioni delle lenti in modo da ottenere una calibrazione che consenta un'eccellente accuratezza nella fase di scansione.
+
To calibrate a camera we could use the following way: focus on a chessboard whose dimensions are known, which is situated on the ground plane. From its examination, it is possible to obtain all the necessary piece of information.
 
+
 
+
Un modo per calibrare la telecamera è quello di inquadrare una scacchiera di dimensioni note collocata sul piano di terra. Dalla sua osservazione è possibile ricavare tutte le informazioni necessarie.
+
  
  
Line 35: Line 32:
 
</div>
 
</div>
  
=== Come capisco se il mio scanner non è calibrato? ===
+
=== How can I understand if my Scanner is not calibrated?===
 
<div align="justify">   
 
<div align="justify">   
Evidenti  disallineamenti sulla superficie nel risultato della scansione sono il principale indicatore di uno scanner non calibrato.
+
The main indicators of a non calibrated Scanner are clear disalignments on the surface, in the scanning process result.
  
Il messaggio di errore ''L'oggetto si è mosso durante la scansione'' può comparire nel caso  di una cattiva calibrazione dello scanner.
 
  
La figura che segue mostra lo stesso oggetto, a sinistra scansionato con uno scanner calibrato perfettamente, a destra con uno scanner non calibrato.
+
In case of a bad calibration of the Scanner, it could be possible to see an error message of this type '''The object has moved during the Scanning process'''.
 +
 
 +
 
 +
The picture here below, shows the same object, on the left we have a scan with a Scanner perfectly calibrated, while on the right a scan with a Scanner non-calibrated.
 +
 
  
  
Line 51: Line 51:
 
</div>
 
</div>
  
=== Ogni quanto è necessario ricalibrare lo scanner? ===
+
=== How many times it is necessary to calibrate the Scanner? ===
 
<div align="justify">   
 
<div align="justify">   
La qualità dei materiali costruttivi e la solidità della struttura  fanno sì che se utilizzato correttamente non sia necessario calibrare frequentemente lo scanner.
+
The quality of the materials and the solidity of the structure permit not to calibrate very often the Scanner if it has used correctly.
 +
 
  
Eccessivi sbalzi di temperatura o urti di una certa entità possono tuttavia  rendere indispensabile la calibrazione della macchina.
+
Nevertheless, excessive jump in the temperature or bad hits, could damage the machine and make the calibration necessary.
  
Piccoli disallineamenti e calibrazioni non perfette non risultano evidenti nel risultato finale, si consiglia perciò di effettuare la calibrazione dello scanner ogni sei mesi.
+
Small disalignments and imperfect calibrations are not evident on the final result, for this reason it is recommended to make the Scanner calibration every six months.
  
 
</div>
 
</div>
  
=== Attivazione del modulo di Calibrazione ===
+
=== Calibration module activation ===
 
<div align="justify">   
 
<div align="justify">   
Per attivare il modulo di calibrazione è necessario richiedere e caricare il file di licenza.
+
In order to activate the calibration module it is necessary to request and to load the license file.
  
Per la richiesta del modulo di calibrazione occorre comunicare il numero seriale che appare accedendo al menù [[image:help.png]] ''Aiuto'' e selezionando [[image:about.png]] ''Circa''.
+
For the calibration module request, it is necessary to communicate the serial number that appears, by login into the menu [[image:help.png]] ''Help'' and by selecting [[image:about.png]] ''About''.
  
Dopo aver ricevuto il file di licenza "eds.calibration.license", sarà necessario  caricarlo accedendo al menù [[image:Eds.File.png]] ''File'' e selezionare  [[image:Activate module.png]] ''Attivare Modulo di calibrazione'', scegliere  la cartella dove si trova il file di licenza e cliccare due volte su di esso. 
 
  
 +
After having received the license "eds.calibration.license", it will be necessary to charge it by login into the menu [[image:Eds.File.png]] ''File'' and by selecting [[image:Activate module.png]] ''Activate calibration module'', choose the folder where there is the license file and tap twice on it.
  
  

Revision as of 17:57, 3 December 2013

Contents

What is calibration?

In a 3D scan system, calibrate a camera means extract the position of the ground plane in the chosen reference and a serial of parameters, necessary to the following elaboration. These parameters are called intrinsics and they are:

  • focal length (fo)
  • image center (cc)
  • radial distortion factor (k)

As told before, the goal of the camera calibration is to estimate the intrinsics parameters (focal length, image center and radial distortion factor) and to identify the position of the ground plane in the solidal reference with the camera.

Cameras calibration is a basic procedure to extract metric piece of information from 2D images. A camera is usually represented by the central projector model pinhole camera. This model is defined by two parameters:

  • Intrinsic parameters (from three to five): they describe the camera independently from their position in the space.
  • Extrinsic parameters (six): they describe camera position in the space, independently from their internal features.

This model must be conveniently correct to consider optical distortions introduced by camera lenses. The use of this kind of lenses, introduces a distortion which is necessary to balance before making a whatever calculation.


In literature there are several models to describe this kind of phenomenon. The model which has been studied and developed by Maestro 3D, estimates the lenses distortions in order to obtain a calibration which allows an excellent attention in the scanning process.


To calibrate a camera we could use the following way: focus on a chessboard whose dimensions are known, which is situated on the ground plane. From its examination, it is possible to obtain all the necessary piece of information.


Maestro3d.calibration.introduction.png


How can I understand if my Scanner is not calibrated?

The main indicators of a non calibrated Scanner are clear disalignments on the surface, in the scanning process result.


In case of a bad calibration of the Scanner, it could be possible to see an error message of this type The object has moved during the Scanning process.


The picture here below, shows the same object, on the left we have a scan with a Scanner perfectly calibrated, while on the right a scan with a Scanner non-calibrated.



Maestro.3D.Easy.Dental.Scan. Calibration.Lower.Arch.Comparison.png


How many times it is necessary to calibrate the Scanner?

The quality of the materials and the solidity of the structure permit not to calibrate very often the Scanner if it has used correctly.


Nevertheless, excessive jump in the temperature or bad hits, could damage the machine and make the calibration necessary.

Small disalignments and imperfect calibrations are not evident on the final result, for this reason it is recommended to make the Scanner calibration every six months.

Calibration module activation

In order to activate the calibration module it is necessary to request and to load the license file.

For the calibration module request, it is necessary to communicate the serial number that appears, by login into the menu Help.png Help and by selecting About.png About.


After having received the license "eds.calibration.license", it will be necessary to charge it by login into the menu Eds.File.png File and by selecting Activate module.png Activate calibration module, choose the folder where there is the license file and tap twice on it.


Kit di Calibrazione

Il kit di calibrazione si compone di due oggetti:

  • un provino di calibrazione per le telecamere che consiste di in un piattino di vetro rettificato sul quale è stata attaccata una stampa in alta definizione di una "scacchiera".


Maestro3d.Easy.Dental.Scan.Camera.Calibration.Kit.png


  • un provino per la calibrazione della tavola rotante ovvero un disco adesivo bianco con 3 anelli neri da applicare sul piattino porta arcata.


Maestro3d.Easy.Dental.Scan.Calibration.Rotary.Table.png

Toolbar Principale

La toolbar principale permette di accedere a tutte le funzionalità di Easy Dental Scan modulo di calibrazione.



Maestro3d.Easy.Dental.Scan.Calibration.Toolbar.png


  • 1) Connessione
  • 2) Disconnessione
  • 3) Snapshot
  • 4) Calcolare la calibrazione dalle immagini acquisite
  • 5) Calibrazione della tavola rotante



Calibrazione delle telecamere - step by step

La procedura di calibrazione delle telecamere consiste nell'acquisizione di un certo numero di immagini, ciascuna delle quali riprende il provino di calibrazione in posizione diversa l'una dall'altra.

Per entrare nel modulo di calibrazione, dalla schermata iniziale di Easy Dental Scan premere il pulsante Eds.calibration.png Calibrazione.

Prima di avviare la vera e propria procedura di calibrazione, posizionare nello scanner il piattino porta arcata con sopra il provino di calibrazione. Per ottenere una "buona" calibrazione è opportuno posizionare il provino con un angolo di circa 45° rispetto al piattino, come sostegno si può usare semplicemente della plastilina come mostrato in figura.


Maestro3D.Easy.Dental.Scan.Calibration.Angle.png



Premendo il pulsante Eds.CalibrationStart.png Connessione sulla barra degli strumenti del modulo di calibrazione, il software procederà alla inizializzazione delle varie componenti dello scanner e sul monitor appariranno due finestre, riferibili alla telecamera sinistra (LX) e destra (RX). Inoltre sulla barra degli strumenti del software si abiliteranno i pulsanti Eds.Calibration exit.png Disconnessione e Eds.Snap.png Snapshot.

Per raggiungere un buon risultato nella calibrazione dello strumento sarà necessario acquisire almeno cinque immagini (consigliate dieci) del provino di calibrazione. Questa fase è molto importante ed è necessario prestare molta attenzione a quanto segue:


  • Verificare prima di ogni snapshot che tutti i quadrati della scacchiera del provino di calibrazione siano illuminati e visibili sia nella telecamera sinistra (LX) che in quella di destra (RX).

La figura che segue mostra il provino correttamente posizionato e acquisito dalle telecamere


Maestro3D.Easy.Dental.Scan.Camera.Calibration.Right.png


La seguente figura invece mostra il provino posizionato parzialmente fuori dal campo visivo delle telecamere e in basso il messaggio di errore nell'acquisizione


Maestro3D.Easy.Dental.Scan.Camera.Calibration.Wrong.png


  • Il provino di calibrazione deve essere in posizione diversa tra un'acquisizione (Snapshot) e l'altra.
  • Sul provino non ci devono essere macchie o graffi che possano impedire il corretto riconoscimento di tutti i quadrati.
  • Il provino di calibrazione deve essere assolutamente immobile al momento dell'acquisizione.
  • Una volta posizionato il provino sul piattino, fare attenzione che non ci siano riflessi sulla superficie. Tali riflessi possono essere causati o dalla sorgente luminosa e in tal caso basterà spostare il provino di calibrazione, o dalla luce ambientale, si consiglia perciò di effettuare le acquisizioni con lo sportello dello scanner chiuso.
  • Onde evitare che sbalzi di temperatura possano vanificare la corretta calibrazione dello scanner si consiglia di effettuare l'operazione di calibrazione nelle stesse condizioni di temperatura nelle quali si andrà ad utilizzare lo strumento.


Per la corretta acquisizione del provino di calibrazione procedere come segue:

  • Posizionare il provino di calibrazione sul piattino porta arcata dello scanner nella posizione più vicina alla sorgente luminosa, chiudere lo sportello dello scanner e premere il pulsante Eds.Snap.png Snapshot. Il software provvederà all'acquisizione dell'oggetto e se tale processo andrà a buon fine, nel centro dello schermo apparirà un numero progressivo in verde e in basso il messaggio ok. Qual'ora l'acquisizione non andasse a buon fine, il progressivo non aumenterà e in basso nello schermo apparirà il messaggio di errore Fallito in rosso.
  • Ripetere la precedente operazione questa volta posizionando il provino nel centro del piattino porta arcata, premere il pulsante Eds.Snap.png Snapshot, e attendere che il software acquisisca l'immagine e il progressivo verde aumenti di una unità.
  • Cambiare posizione al provino e posizionarlo nel punto più distante dalla sorgente luminosa, premere il pulsante Eds.Snap.png Snapshot e attendere la conferma dell'avvenuta acquisizione.
  • Procedere allo spostamento del provino nella posizione di estrema destra e sinistra rispetto alla sorgente luminosa ed effettuare l'acquisizione delle immagini con il pulsante Eds.Snap.png Snapshot.



Maestro3D.Easy.Dental.Scan.Camera.Calibration.png



La procedura descritta garantisce che l'intera area di copertura delle telecamere venga utilizzata. Si consiglia tuttavia di ripetere le operazioni di acquisizione per altre cinque volte, per un totale di 10 snapshot, facendo attenzione a cambiare ogni volta la posizione del provino.

Dopo aver acquisito con successo almeno cinque immagini, sulla barra degli strumenti del modulo di calibrazione si abiliterà il pulsante Eds.CalibrationCompute.png Calcolare la calibrazione dalle immagini acquisite. Premendolo il software disabilita tutti i pulsanti sulla barra degli strumenti, disconnette le periferiche e procede ad elaborare i dati acquisiti. Se la procedura va a buon fine apparirà una finestra con il messaggio Info32.png La calibrazione è terminata con successo ed è stata installata. Il file di calibrazione camera.calibration.xml viene automaticamente salvato in Easy Dental Scan.

Premendo il pulsante OK il software tornerà alla schermata iniziale del modulo di calibrazione.

Importante:

  • Il provino di calibrazione dev'essere assolutamente immobile al momento dell'acquisizione.
  • Non effettuare più di una acquisizione con il provino di calibrazione nella stessa posizione.
  • Controllare che la scacchiera del provino di calibrazione sia ripreso completamente da entrambe le telecamere.


Calibrazione della tavola rotante - step by step

Prima di effettuare la procedura di calibrazione della tavola rotante, applicare sul piattino porta arcata il provino come mostrato in figura.


Maestro3d.Easy.Dental.Scan.Calibration.Rotary.Table.Stiker.png


Per accedere alla procedura di calibrazione della tavola rotante cliccare sul pulsante Eds.CalibrationRotaryTableStart.png Calibrazione della tavola rotante posizionato sulla barra degli strumenti nella schermata iniziale del modulo di calibrazione. Posizionare il provino di calibrazione della tavola rotante come mostrato nella figura a sinistra nella schermata, quindi procedere cliccando il pulsante Next.png Avanti in basso a destra.

Lo scanner effettuerà una scansione del provino e il software mostrerà dopo pochi secondi il risultato nella schermata successiva.

Tenendo premuto il Tasto SHIFT sulla tastiera, procedere a marcare il centro degli anelli neri con un doppio click del tasto sinistro del mouse, nell'ordine suggerito nella figura a sinistra.


Maestro3D.Easy .Dental.Scan.rotary.Table.Markers.png


Eventuali modifiche nella posizione o nell'ordine dei punti possono essere effettuate con il pulsante Polyline.delete.png Pulisci oppure Undo.png Annulla.

Dopo aver marcato il centro dei tre anelli si abiliterà in basso a destra il pulsante Next.png Avanti, premendolo il software procederà ad installare il file rotary.table.calibration.xml in Easy Dental Scan.

Personal tools
Namespaces

Variants
Actions
Navigation
Toolbox